Accuracy of residual respiratory event detection by CPAPs: a meta-analysis

verfasst von: Imran H. Iftikhar, Ahmed BaHammam, Haitham Jahrami, Octavian Ioachimescu

Erschienen in: Sleep and Breathing | Ausgabe 5/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Most continuous positive airway pressure (CPAP) machines have built-in manufacturer-specific proprietary algorithms for automatic respiratory event detection (AED) based on very specific respiratory events scoring criteria. With regards to the accuracy of these data from CPAP machines, evidence from the literature seems conflicting, which formed the basis for this meta-analysis.

Methods

A meta-analysis was performed on studies that reported Bland-Altman analysis data on agreement (mean bias and limits of agreement [LoA]) of CPAP-determined apnea-hypopnea index (AHI) at therapeutic pressures (AHIFLOW) with that determined from simultaneously conducted polysomnograms (AHIPSG).

Results

In six studies, ResMed CPAPs were used, and in another six studies, Respironics CPAPs were used, while only one study used Fisher & Paykel (F&P) CPAPs. The pooled mean AHI bias from ResMed CPAP studies was − 1.01 with pooled LoAs from − 3.55 to 1.54 (I2 = 17.5%), and from Respironics CPAP studies, pooled mean AHI bias was − 0.59 with pooled LoAs from − 3.22 to 2.05 (I2 = 0%). Pooled percentage errors (corresponding to LoAs) from four ResMed CPAP studies, four Respironics CPAP studies, and the F&P CPAP study were 73%, 59%, and 112%, respectively. A review of the literature for this meta-analysis also revealed lack of uniformity not only in the CPAP manufacturers’ respiratory events scoring criteria but also in that used for PSGs across the studies analyzed.

Conclusions

Even though the pooled results of mean AHI bias suggest good clinical agreement between AHIPSG and AHIFLOW, percentage errors calculated in this meta-analysis indicate the possibility of a significant degree of imprecision in the estimation of AHIFLOW by CPAP machines.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
3.
Zurück zum Zitat Baek JJJ, Lee S (2016) Accuracy of the auto scoring by the S9 CPAP in patients with obstructive sleep apnea. Sleep Med Res 7(1):26–32CrossRef Baek JJJ, Lee S (2016) Accuracy of the auto scoring by the S9 CPAP in patients with obstructive sleep apnea. Sleep Med Res 7(1):26–32CrossRef
5.
Zurück zum Zitat Ueno K, Kasai T, Brewer G, Takaya H, Maeno K, Kasagi S, Kawana F, Ishiwata S, Narui K (2010) Evaluation of the apnea-hypopnea index determined by the S8 auto-CPAP, a continuous positive airway pressure device, in patients with obstructive sleep apnea-hypopnea syndrome. J Clin Sleep Med 6(2):146–151CrossRefPubMedPubMedCentral Ueno K, Kasai T, Brewer G, Takaya H, Maeno K, Kasagi S, Kawana F, Ishiwata S, Narui K (2010) Evaluation of the apnea-hypopnea index determined by the S8 auto-CPAP, a continuous positive airway pressure device, in patients with obstructive sleep apnea-hypopnea syndrome. J Clin Sleep Med 6(2):146–151CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, Shamseer L, Tetzlaff JM, Akl EA, Brennan SE, Chou R, Glanville J, Grimshaw JM, Hrobjartsson A, Lalu MM, Li T, Loder EW, Mayo-Wilson E, McDonald S, McGuinness LA, Stewart LA, Thomas J, Tricco AC, Welch VA, Whiting P, Moher D (2021) The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. Rev Esp Cardiol (Engl Ed) 74(9):790–799. https://doi.org/10.1016/j.rec.2021.07.010CrossRefPubMed Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, Shamseer L, Tetzlaff JM, Akl EA, Brennan SE, Chou R, Glanville J, Grimshaw JM, Hrobjartsson A, Lalu MM, Li T, Loder EW, Mayo-Wilson E, McDonald S, McGuinness LA, Stewart LA, Thomas J, Tricco AC, Welch VA, Whiting P, Moher D (2021) The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. Rev Esp Cardiol (Engl Ed) 74(9):790–799. https://​doi.​org/​10.​1016/​j.​rec.​2021.​07.​010CrossRefPubMed
10.
Zurück zum Zitat Joosten A, Desebbe O, Suehiro K, Murphy LS, Essiet M, Alexander B, Fischer MO, Barvais L, Van Obbergh L, Maucort-Boulch D, Cannesson M (2017) Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: a systematic review and meta-analysisdagger. Br J Anaesth 118(3):298–310. https://doi.org/10.1093/bja/aew461CrossRefPubMed Joosten A, Desebbe O, Suehiro K, Murphy LS, Essiet M, Alexander B, Fischer MO, Barvais L, Van Obbergh L, Maucort-Boulch D, Cannesson M (2017) Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: a systematic review and meta-analysisdagger. Br J Anaesth 118(3):298–310. https://​doi.​org/​10.​1093/​bja/​aew461CrossRefPubMed
14.
Zurück zum Zitat Gagnadoux F, Pevernagie D, Jennum P, Lon N, Loiodice C, Tamisier R, van Mierlo P, Trzepizur W, Neddermann M, Machleit A, Jasko J, Pepin JL (2017) Validation of the system One RemStar Auto A-Flex for obstructive sleep apnea treatment and detection of residual apnea-hypopnea index: a European randomized trial. J Clin Sleep Med 13(2):283–290. https://doi.org/10.5664/jcsm.6464CrossRefPubMedPubMedCentral Gagnadoux F, Pevernagie D, Jennum P, Lon N, Loiodice C, Tamisier R, van Mierlo P, Trzepizur W, Neddermann M, Machleit A, Jasko J, Pepin JL (2017) Validation of the system One RemStar Auto A-Flex for obstructive sleep apnea treatment and detection of residual apnea-hypopnea index: a European randomized trial. J Clin Sleep Med 13(2):283–290. https://​doi.​org/​10.​5664/​jcsm.​6464CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Schwab RJ, Badr SM, Epstein LJ, Gay PC, Gozal D, Kohler M, Levy P, Malhotra A, Phillips BA, Rosen IM, Strohl KP, Strollo PJ, Weaver EM, Weaver TE, Systems ATSSoCAT (2013) An official American Thoracic Society statement: continuous positive airway pressure adherence tracking systems. The optimal monitoring strategies and outcome measures in adults. Am J Respir Crit Care Med 188(5):613–620. https://doi.org/10.1164/rccm.201307-1282STCrossRefPubMedPubMedCentral Schwab RJ, Badr SM, Epstein LJ, Gay PC, Gozal D, Kohler M, Levy P, Malhotra A, Phillips BA, Rosen IM, Strohl KP, Strollo PJ, Weaver EM, Weaver TE, Systems ATSSoCAT (2013) An official American Thoracic Society statement: continuous positive airway pressure adherence tracking systems. The optimal monitoring strategies and outcome measures in adults. Am J Respir Crit Care Med 188(5):613–620. https://​doi.​org/​10.​1164/​rccm.​201307-1282STCrossRefPubMedPubMedCentral
Metadaten
Titel
Accuracy of residual respiratory event detection by CPAPs: a meta-analysis
verfasst von
Imran H. Iftikhar
Ahmed BaHammam
Haitham Jahrami
Octavian Ioachimescu
Publikationsdatum
30.01.2023
Verlag
Springer International Publishing
Erschienen in
Sleep and Breathing / Ausgabe 5/2023
Print ISSN: 1520-9512
Elektronische ISSN: 1522-1709
DOI
https://doi.org/10.1007/s11325-023-02780-w

Weitere Artikel der Ausgabe 5/2023

Sleep and Breathing 5/2023 Zur Ausgabe

Sleep Breathing Physiology and Disorders • Original Article

Contribution of central sleep apnea to severe sleep apnea hypopnea syndrome

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Bei seelischem Stress sind Checkpoint-Hemmer weniger wirksam

03.06.2024 NSCLC Nachrichten

Wie stark Menschen mit fortgeschrittenem NSCLC von einer Therapie mit Immun-Checkpoint-Hemmern profitieren, hängt offenbar auch davon ab, wie sehr die Diagnose ihre psychische Verfassung erschüttert

Antikörper mobilisiert Neutrophile gegen Krebs

03.06.2024 Onkologische Immuntherapie Nachrichten

Ein bispezifischer Antikörper formiert gezielt eine Armee neutrophiler Granulozyten gegen Krebszellen. An den Antikörper gekoppeltes TNF-alpha soll die Zellen zudem tief in solide Tumoren hineinführen.

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.