Change your Angle of View

Sinusoidal C-Arm Movement in Cranial Flat-panel CT to Improve Image Quality

verfasst von: Hannes Luecking, Philip Hoelter, Stefan Lang, Manuel Schmidt, Felix Eisenhut, Arnd Doerfler

Erschienen in: Clinical Neuroradiology | Ausgabe 4/2022

Abstract

Background

Artifacts from surrounding bony structures, especially from the petrous bones, regularly impair soft tissue computed tomography (CT) imaging of the middle and posterior fossa. This affects flat-panel CT in particular. Sinusoidal movement of the C‑arm during acquisition (i.e. craniocaudal tilting along with semicircular rotation) is supposed to reduce artifacts, thus enhancing soft tissue imaging quality.

Methods

In the work-up of ischemic stroke or subarachnoid hemorrhage 40 patients underwent multi-slice CT (MS-CT) and either plain circular (cFP-CT; n = 20) or sinusoidal (sFP-CT; n = 20) flat-panel CT within a short interval. Two independent readers analyzed MS-CT and FP-CT datasets for recognizability of eight different brain structures and three typical types of artifacts according to a predetermined score.

Results

Interrater reliability was moderate for cFP-CT (κ = 0.575) and good to very good for ratings of MS-CT and sFP-CT (κ = 0.651 to κ = 1). MS-CT was rated to be significantly better than cFP-CT and sFP-CT (p < 0.0001) in the overall score. Yet, sFP-CT was rated to be significantly superior to cFP-CT (overall p < 0.0001) regarding most anatomical regions and petrous bone artifacts.

Conclusion

Compared to a standard circular protocol, sinusoidal C‑arm movement in cranial FP-CT can significantly reduce artifacts in the posterior fossa and, moreover, can improve visualization of most supratentorial and infratentorial anatomical structures.

Introduction

Flat-panel computed tomography (FP-CT) has been available for many years in the surrounding of cerebral digital subtraction angiography (DSA), e.g. for therapy planning [1, 2]. Contrast-enhanced 3D datasets can be beneficial for the work-up of several kinds of pathologies from simple cerebral aneurysms to complex arteriovenous malformations.
Non-enhanced FP-CT can be used to replace multi-slice cranial CT (MS-CT) in certain settings. It allows for MS-CT-like imaging without having to move patients to the CT suite, thus reducing transportation time. Typical settings could involve re-evaluation of patients with acute stroke after relocation to a thrombectomy center or verification of proper placement of ventricular drainages in patients with subarachnoid or intraventricular hemorrhage prior to DSA (one-stop-shopping).
Although image quality has improved for both modalities over the years [3], FP-CT performs still behind MS-CT regarding soft tissue contrast and bony hardening artifacts. This affects the posterior fossa in particular, especially the brainstem.
In hitherto FP-CT the C‑arm follows a plain (semi)circular path. Sinusoidal movement adds a craniocaudal component of motion along this circular path that allows for varying angels of view to the volume of interest. Fig. 1 illustrates the trajectory of the C‑arm along its circular and craniocaudal path. This is supposed to reduce the impact of hardening artifacts from adjacent bony structures, thus improving soft tissue imaging quality.
This evaluation compares MS-CT to FP-CT runs of the respective patients, 2 groups of each 20 patients (40 patients in total) underwent either MS-CT and circular FP-CT (cFP-CT) or MS-CT and sinusoidal FP-CT (sFP-CT).

Methods

Inclusion

From our database a total of 40 patients were selected who had undergone cFP-CT and MS-CT (n = 20) or sFP-CT and MS-CT (n = 20) within 3 h prior to FP-CT. Whereas MS-CT is done as first line imaging for most patients arriving via the emergency ward, FP-CT is performed regularly in our center in stroke and SAH work-up to rule out procedure-related complications after neuroendovascular interventions. Datasets with major foreign objects, e.g. large coil packages, implanted in between MS-CT and FP-CT, were excluded in order to obtain good comparability of soft tissue contrast. Additional comparison of FP-CT to MS-CT as well as comparison of both MS-CT subgroups among each other were chosen to ensure general comparability of both FP-CT subgroups (rule-out of casual bias in FP-CT subgroup allocation).
SFP-CT is set as system standard in our routine setup of the angiographic system. For collection of cFP-CT datasets, the system standard was changed from sFP-CT to cFP-CT for a time span of 3 months.

CT-Imaging

MS-CT was done on a 128-row multi-slice scanner Somatom Definitions AS+ (Siemens Healthineers, Forchheim, Germany). Secondary reconstructions were drawn for the analysis with 4.8 mm slice thickness and 4.8 mm spacing. Predetermined center value was 40 HU and window width was 100 HU. Changing center and window settings was allowed during assessment.

FP-CT Imaging

Native 8s circular FP-CT and 7s sinusoidal FP-CT runs were performed on an ICONO biplane angiographic system (Siemens Healthineers, Forchheim, Germany), with both protocols being factory standard. In analogy to the MS-CT reconstruction protocol, 4.8 mm reconstructions were done with manual setting of center and window values. Again, changing center and window settings during assessment was allowed.
SFP-CT acquires 100 more projections than cFP-CT (596 vs. 496). This results in a coverage of 220° in sFP-CT rather than 200° in cFP-CT. Furthermore, sFP-CT uses 4 × 4 binning instead of 2 × 2 binning in cFP-CT. Binning describes the summation of the signal of several detector units in electronic post-processing, with binning of more units resulting in a smoother image. For further scan parameters of MS-CT and FP-CT see Table 1.
Table 1
Scan parameters for circular and sinusoidal protocols
 
Voltage
Angle
Stepping
Projections
Binninga
Dose
8s circular
109 kV
200°
0.04°/frame
496
2 × 2
1.820 μGy/frame
7s sinusoidal
109 kV
220°
0.04°/frame
596
4 × 4
1.820 μGy/frame
a Pooling of several detector elements (higher binning results in lower spatial resolution and lower noise and vice versa)

Imaging Evaluation and Statistical Analysis

All 80 datasets were evaluated independently by two readers with more than 10 years of experience in neuroimaging. A simple scoring system was created to comprehensively assess brain soft tissue, bony and foreign object hardening artifacts. Sinusoidal C‑arm movement is mainly supposed to reduce petrous bone artifacts in the posterior fossa. Besides assessing the brainstem and cerebellum, also supratentorial structures were included in the scoring system in order to detect whether the technique could also provide advantages in other anatomical regions. This led to a catalogue of eight anatomic structures and another three types of artifacts being assessed, each with a threefold selection, resulting in a maximum score of 33 points (Table 2).
Table 2
Assessment results for both raters (mean value through all patients) using Mann-Whitney-U-Test. Last column gives significance level for comparison between 8s circular and 7s sinusoidal runs
 
Rater 1 & 2
Criterion
MS-CT
8s circular
MS-CT
7s sine spin
p value 8s/7s
GWMa frontalb
3
2.6
3
2.8
p = 0.042
GWMa parietalb
3
2.4
3
2.7
p = 0.012
GWMa temporalb
3
1.8
2.9
1.9
p = 0.164
GWMa occipitalb
3
1.7
3
2.1
p = 0.012
Insular cortexb
2.9
1.8
3
2
p = 0.142
Basal gangliab
2.9
2.1
3
2.5
p = 0.003
Brainstemb
2.5
1.2
2.6
1.8
p < 0.0001
Cerebellumb
3
1.2
3
1.5
p = 0.002
Petrous bone artifactsc
2.7
1.4
3
1.9
p < 0.0001
Skull hardening artifactsc
3
2.3
3
2.4
p = 0.169
Foreign objects artifactsc
3
3
3
2.9
p = 0.352
Sum
31.8
21.2
32.3
24.4
P < 0.0001
a Grey/white matter contrast
b Good (3 points), fair (2 point), poor (1 point)
c Mild (3 points), mediocre (2 points), severe (1 point)
Also, the recognizability of pathologies correlated with the actual cause of admission was recorded but was not part of the scoring system. Regarding tissue assessment there were 3 grades (good: 3 points; fair: 2 points; poor: 1 point) while for artifact assessment the grading was counted contrariwise (no/minor artifacts: 3 points; 2 mediocre artifacts: 2 points; severe artifacts: 1 point).
Blinding the readers to the type of imaging (MS-CT vs. FP-CT) was not appropriate as the nature of the respective reconstruction is obvious to an experienced reader (see Fig. 2).
SPSS 28 (IBM, Armonk, NY, USA) was used for statistical analysis. Ordinal values were checked for normal distribution using Kolmogorov-Smirnov test. Due to the mostly non-normal distribution, Mann-Whitney U‑test was used to compare subgroups. Inter-rater correlation was measured using Cohen’s kappa κ.

Results

Patients were admitted for either ischemic stroke (n = 16 for the cFP-CT group; n = 18 for the sFP-CT group) or for subarachnoid hemorrhage (n = 4 vs. n = 2) and, thus underwent MS-CT immediately after admission and FP-CT in the course of thrombectomy or SAH-/aneurysm work-up.
FP-CT was consistently performed with the patient under general anesthesia in the treatment setting of their respective primary disease in all patients, ruling out motion artifacts.

Assessment Rating

Both MS-CT groups (circular control and sinusoidal control) were not found to differ significantly regarding overall assessment score by both readers (p = 0.100; 95% CI), indicating that both the circular and sinusoidal groups had no major bias (e.g. foreign material) systematically compromising multi-slice and flat-panel imaging. Table 2 provides combined results of both readers for all assessed anatomical structures and artifacts. Separate results of both raters are provided in supplementary table S2. SFP-CT was rated significantly superior to cFP-CT in assessment of grey/white matter contrast of the frontal, parietal and occipital lobes, and visualization of basal ganglia, brainstem and cerebellum. Petrous bone artifacts were found to be significantly less in sFP-CT runs. No significant difference was found for the temporal lobe and insular cortex and skull and foreign objects hardening artifacts.
Both sFP-CT (mean score = 24.5; p < 0.0001) and cFP-CT (mean score = 21.3; p < 0.0001) runs were found to perform significantly lower than MS-CT (mean score = 32.0) by both raters with respect to the overall score.

Interrater Reliability

Cohen’s kappa was calculated to check interrater reliability in assessment of visibility of anatomical structures and artifacts. Each subset (cFP-CT, sFP-CT, circular control and sinusoidal control) was calculated separately. Results are provided in Table 3. According to the interpretation of Altman, interrater reliability was good (κ ≥ 0.60) or very good (κ ≥ 0.80) regarding assessment of anatomical structures and artifacts in all subgroups, except for anatomical structures in cFP-CT, where reliability was found to be moderate (κ ≥ 0.40).
Table 3
Interrater (IR) reliability for assessment of anatomical structures and artifacts for separate modalities
 
IR reliability anatomy
IR reliability artifacts
8s multi-slice control
κ = 0.651
κ = 0.621
8s circular FP-CT
κ = 0.575
κ = 0.821
7s multi-slice control
κ = 0.656
κ = 1
7s sinusoidal FP-CT
κ = 0.842
κ = 0.757
Both raters found a higher number of pathologic findings in the cFP-CT and sFP-CT datasets compared to their respective MS-CT groups (rater 1: MS-CT:cFP-CT=15:19; MS-CT:sFP-CT=11:22; rater 2: MS-CT:cFP-CT=16:17; MS-CT:sFP-CT=11:22). This demonstrates the general ability of FP-CT of either kind to detect and assess admission-related pathologies. Yet, given the time interval of mean 146 ± 30.2 min (range 63–189 min) between MS-CT and FP-CT, it lies in the nature especially of ischemic stroke to show progressive visibility in the course of time (Fig. 3).

Radiation Dose

The dose area product (DAP) was significantly higher in sFP-CT (mean 5076 µGym2) than in cFP-CT (mean 4583 µGym2), p = 0.0021.

Discussion

Non-enhanced flat panel CT of the brain as an alternative for “conventional” multi-slice CT can be used for multiple purposes in the surrounding of neuroendovascular imaging and treatment. Imaging quality, yet, is inferior to MS-CT regarding soft tissue contrast and susceptibility to bone hardening artifacts, which affects the posterior fossa in particular. Sinusoidal C‑arm movement is supposed to reduce petrous bone and hardening artifacts in general, by providing different angles of view not only along a circular path but also in the craniocaudal direction.
It was the aim of this analysis to compare image quality of cFP-CT and sFP-CT among each other and to common MS-CT.

Imaging Characteristics

For reading of MS-CT, a standard center/window setting of 40/100 Hounsfield units is used, which is suitable in most cases. FP-CT, on the other hand, was found to require distinctive changes to center and window settings to assess different areas, e.g. cerebral cortex, basal ganglia or middle and posterior fossa, in all datasets. This increases assessment effort and could, depending on the reader’s experience and dedication, lead to a more reader-dependent quality of reporting in FP-CT.
With less artifacts especially in the posterior fossa, effort to find appropriate window settings is, by subjective perception, slightly reduced in the sFP-CT group (see Fig. 2, row 2), but the basic statement remains true for both groups of FP-CT.
SFP-CT is reconstructed with a 4 × 4 binning of detector elements, as opposed to 2 × 2 binning in cFP-CT. As there are no factory standard protocols available that match regarding this setting, it cannot be excluded that the higher binning in sFD-CT could, by reduction of noise, also contribute to the more favorable results of sFP-CT. Setting up a respective sFP-CT protocol with 2 × 2 binning required major changes to the system including recalibration and consistency check and, thus, was not meaningful for the evaluation.

Interrater Reliability

Although both raters have long experience in assessing FP-CT and interrater reliability showed mostly “good” and “very good” reliability, there was only “moderate” reliability for cFP-CT. This indicates that the individual and subjective perception of imaging quality may differ relevantly, especially if the quality is below gold standard (MS-CT). Pathological findings of clinical relevance, yet, can be depicted reliably by FP-CT, as shown by other authors [4]. Interrater reliability regarding assessment of anatomical structures was best for sFP-CT. This raises the assumption that improvement of imaging characteristics and quality as provided with sFP-CT can contribute to making reading of FP-CT more reliable, potentially enabling readers to gather information more easily.

Assessment Results

With an average of 32/33 points (97%) MS-CT was rated significantly better regarding assessment of anatomical structures and susceptibility to artifacts compared to sFP-CT (24.5/33 points, 74%) and cFP-CT (21.3/33 points, 65%), (p < 0.0001, 95% CI). Yet, sFP-CT was found to be significantly better than cFP-CT (p < 0.0001) in the overall score and in most anatomical areas except the temporal lobe and insular cortex. Gray/white matter contrast in the frontal, occipital and parietal lobes was visualized significantly better (Fig. 2, right column), possibly allowing the visual or automated determination of aspects scores in one-stop-shopping or post thrombectomy scans [57]. Petrous bone artifacts, which are one key point when it comes to image quality in the posterior fossa, were rated to be significantly lower in sFP-CT (p < 0.0001), going along with a, again, significantly better visualization of the brainstem and cerebellum.
Although visualization of the temporal lobe and insula was rated slightly better in sFP-CT than in cFP-CT, the result was not significant. The authors assume that this is caused by scatter radiation from the skull base and the lower soft tissue resolution of FP-CT, which affects cFP-CT and sFP-CT similarly.
If admission-related pathologies were discovered, this was also noted. Due to the natural course of ischemic brain lesions, FP-CT revealed more pathologies than MS-CT, which consistently was performed prior to FP-CT in the course of diagnostics and treatment. Although not assessed in particular, this indicates that FP-CT—regardless of scan parameters—can be able to detect relevant pathologies, such as hemorrhage or subacute infarcts.

Radiation Dose

The dose-area product (DAP) that is automatically calculated and displayed by the angiographic system, was approximately 10.7% (mean ~500 µGym2) higher in sFP-CT than in cFP-CT. Sinusoidal scans do not allow for craniocaudal collimation of the beam to make sure that the whole volume of interest is covered along the trajectory of the C‑arm. Furthermore, a coverage of 220° in sFP-CT rather than 200° in cFP-CT adds to an increase in DAP. The lack of possible collimation may lead to the assumption that a relevant share of the additional radiation probably does not interfere with the patient. Yet, these peculiarities of sFP-CT should be kept in mind as, for instance, eye lenses cannot to be spared by proper bedding of the patient’s head [810].

Limitations

The study is limited by the retrospective design and limited number of patients. Comparison of effective radiation dose would require elaborate phantom tests.

Conclusion

Sinusoidal C‑arm movement in flat-panel imaging as new feature of recent angiographic systems provides significantly superior image quality of supratentorial and infratentorial brain tissue compared to regular “circular” flat-panel CT. Petrous bone artifacts are significantly reduced.

Conflict of interest

H. Luecking, P. Hoelter, S. Lang, M. Schmidt, F. Eisenhut and A. Doerfler declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Neurologie

Kombi-Abonnement

Mit e.Med Neurologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes, den Premium-Inhalten der neurologischen Fachzeitschriften, inklusive einer gedruckten Neurologie-Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Hausegger KA, Fürstner M, Hauser M, Smetana F, Kau T. Klinische Anwendung der Flachdetektor-CT im Angio-OP [Clinical application of flat-panel CT in the angio suite]. Rofo. 2011;183:1116-22.CrossRef Hausegger KA, Fürstner M, Hauser M, Smetana F, Kau T. Klinische Anwendung der Flachdetektor-CT im Angio-OP [Clinical application of flat-panel CT in the angio suite]. Rofo. 2011;183:1116-22.CrossRef
2.
Zurück zum Zitat Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A. The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol. 2006;27:330-2. Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A. The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol. 2006;27:330-2.
3.
Zurück zum Zitat Fahrig R, Jaffray DA, Sechopoulos I, Webster Stayman J. Flat-panel conebeam CT in the clinic: history and current state. J Med Imaging (Bellingham). 2021;8:052115. Fahrig R, Jaffray DA, Sechopoulos I, Webster Stayman J. Flat-panel conebeam CT in the clinic: history and current state. J Med Imaging (Bellingham). 2021;8:052115.
4.
Zurück zum Zitat Struffert T, Eyupoglu IY, Huttner HB, Engelhorn T, Doelken M, Saake M, Ganslandt O, Doerfler A. Clinical evaluation of flat-panel detector compared with multislice computed tomography in 65 patients with acute intracranial hemorrhage: initial results. Clinical article. J Neurosurg. 2010;113:901-7.CrossRef Struffert T, Eyupoglu IY, Huttner HB, Engelhorn T, Doelken M, Saake M, Ganslandt O, Doerfler A. Clinical evaluation of flat-panel detector compared with multislice computed tomography in 65 patients with acute intracranial hemorrhage: initial results. Clinical article. J Neurosurg. 2010;113:901-7.CrossRef
5.
Zurück zum Zitat Doerfler A, Gölitz P, Engelhorn T, Kloska S, Struffert T. Flat-Panel Computed Tomography (DYNA-CT) in Neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke. Clin Neuroradiol. 2015;25 Suppl 2:291-7.CrossRef Doerfler A, Gölitz P, Engelhorn T, Kloska S, Struffert T. Flat-Panel Computed Tomography (DYNA-CT) in Neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke. Clin Neuroradiol. 2015;25 Suppl 2:291-7.CrossRef
6.
Zurück zum Zitat Bouslama M, Haussen DC, Grossberg JA, Barreira CM, Bom IMJV, Nijnatten FV, Grünhagen T, Moyer L, Frankel MR, Nogueira RG. Flat-panel detector CT assessment in stroke to reduce times to intra-arterial treatment: A study of multiphase computed tomography angiography in the angiography suite to bypass conventional imaging. Int J Stroke. 2021;16:63-72.CrossRef Bouslama M, Haussen DC, Grossberg JA, Barreira CM, Bom IMJV, Nijnatten FV, Grünhagen T, Moyer L, Frankel MR, Nogueira RG. Flat-panel detector CT assessment in stroke to reduce times to intra-arterial treatment: A study of multiphase computed tomography angiography in the angiography suite to bypass conventional imaging. Int J Stroke. 2021;16:63-72.CrossRef
7.
Zurück zum Zitat Pfaff JAR, Schönenberger S, Herweh C, Ulfert C, Nagel S, Ringleb PA, Bendszus M, Möhlenbruch MA. Direct Transfer to Angio-Suite Versus Computed Tomography-Transit in Patients Receiving Mechanical Thrombectomy: A Randomized Trial. Stroke. 2020;51:2630-8. Erratum in: Stroke. 2020;51:e346.CrossRef Pfaff JAR, Schönenberger S, Herweh C, Ulfert C, Nagel S, Ringleb PA, Bendszus M, Möhlenbruch MA. Direct Transfer to Angio-Suite Versus Computed Tomography-Transit in Patients Receiving Mechanical Thrombectomy: A Randomized Trial. Stroke. 2020;51:2630-8. Erratum in: Stroke. 2020;51:e346.CrossRef
8.
Zurück zum Zitat Struffert T, Lang S, Scholz R, Hauer M, Dörfler A. Strahlendosis bei zerebraler Angiographie und Flachdetektor-CT-Applikationen in der Neuroradiologie [Radiation dose in cerebral angiography and flat detector CT applications in neuroradiology]. Radiologe. 2015;55:654-62.CrossRef Struffert T, Lang S, Scholz R, Hauer M, Dörfler A. Strahlendosis bei zerebraler Angiographie und Flachdetektor-CT-Applikationen in der Neuroradiologie [Radiation dose in cerebral angiography and flat detector CT applications in neuroradiology]. Radiologe. 2015;55:654-62.CrossRef
9.
Zurück zum Zitat Kyriakou Y, Richter G, Dörfler A, Kalender WA. Neuroradiologic applications with routine C-arm flat panel detector CT: evaluation of patient dose measurements. AJNR Am J Neuroradiol. 2008;29:1930-6.CrossRef Kyriakou Y, Richter G, Dörfler A, Kalender WA. Neuroradiologic applications with routine C-arm flat panel detector CT: evaluation of patient dose measurements. AJNR Am J Neuroradiol. 2008;29:1930-6.CrossRef
10.
Zurück zum Zitat Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A. Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol. 2014;24:1257-65.CrossRef Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A. Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol. 2014;24:1257-65.CrossRef
Metadaten
Titel
Change your Angle of View
Sinusoidal C-Arm Movement in Cranial Flat-panel CT to Improve Image Quality
verfasst von
Hannes Luecking
Philip Hoelter
Stefan Lang
Manuel Schmidt
Felix Eisenhut
Arnd Doerfler
Publikationsdatum
05.05.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Neuroradiology / Ausgabe 4/2022
Print ISSN: 1869-1439
Elektronische ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-022-01172-z

Weitere Artikel der Ausgabe 4/2022

Clinical Neuroradiology 4/2022 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.